A Quantitative Cleanability Index for Implant-Supported Prostheses: Geometric Definition, Computational Method, and a Protocol for Validation Against Residual Plaque
- Paper type
- Methodological development with measurement validation protocol
- Status
- Original Work · Working Paper
Contents
Abstract
Background. Plaque control is the basis of preventing peri-implant disease, and prostheses that obstruct hygiene are associated with peri-implantitis. In one clinical series, 74% of implants were judged to have no access for proper oral hygiene, and such implants accounted for a high proportion of peri-implantitis diagnoses. In vitro studies show that interdental aids remove artificial biofilm incompletely and that removal depends on crown contour. Yet access remains a clinical judgment recorded as present or absent. A review of the supracrestal complex concluded that evidence on restricted access is limited, and no geometric definition of access was found in the searches run for this manuscript.
Objective. To define a cleanability index computed from the digital geometry of a prosthesis and the adjacent mucosa, to describe a method for computing it, and to specify a protocol for validating it against residual plaque.
Framework. Each cleaning instrument is represented by a geometric model with a rigid core, a compliant working envelope, and a set of feasible approach directions. A surface point is reachable if some collision-free pose places the working envelope on it. The index is the reachable fraction of a defined critical surface, for a single instrument or a kit. A linear model links the index to residual plaque through a removal efficiency on reached surface.
Proposed methods. Instrument models are parameterized from bench measurements. Stage 1 compares predicted reach maps with three-dimensional maps of residual artificial biofilm after standardized cleaning of printed prostheses. Stage 2 is a cross-sectional clinical study relating the index to plaque on retrieved screw-retained prostheses and to peri-implant bleeding. Reliability, agreement, and construct validity are assessed with prespecified criteria.
Central hypothesis. The index explains residual plaque better than emergence angle, profile category, or binary clinical judgment of access, and unreached surface retains plaque regardless of instrument or operator.
Conclusions. Closed-form cases illustrate how the index behaves for filaments in concavities and for brushes beneath full-arch prostheses. The index has not been computed on clinical geometries, and no validation data exist. Its relation to disease is a hypothesis.
Keywords: implant-supported prosthesis; oral hygiene access; cleanability; peri-implantitis; interdental brush; computer-aided design; accessibility analysis; measurement validation
1. Introduction
Peri-implant mucositis is caused by plaque and resolves when plaque is removed [1–3]. Peri-implantitis develops from it in susceptible sites and is far harder to treat [4, 5]. Prevention therefore depends on plaque control by the patient and on supportive care [6, 7]. In a five-year follow-up of patients with mucositis, peri-implantitis developed in 43.9% of those without preventive maintenance and in 18.0% of those with it [8], and a meta-analysis found that maintenance therapy reduces the occurrence of peri-implant diseases [9].
A patient can remove plaque only from surfaces that an instrument can touch. Implant prostheses differ from teeth in ways that make this harder. The restoration emerges from a narrow platform and widens over a few millimeters, so that its apical contour often faces the mucosa. Splinted units close the embrasures. Full-arch prostheses cover the ridge with a continuous intaglio surface that can be reached only from the sides. The observation that such features matter is old. Serino and Ström examined 23 patients referred for peri-implantitis and reported that 74% of 109 implants had no accessibility for proper oral hygiene; a high proportion of implants with peri-implantitis lacked access (48%), whereas access was rarely associated with the diagnosis (4%) [10]. Later cross-sectional studies linked emergence angles above 30 degrees and convex profiles to peri-implantitis at bone-level implants [11, 12], and reviews now list prosthesis design among local predisposing factors [13–15].
What these studies call access was judged by a clinician at the chair, or inferred from a two-dimensional angle on a radiograph. Neither is a measurement of whether an instrument can reach a surface. A review of the design of the implant supracrestal complex concluded that only limited evidence implicates restricted accessibility for hygiene, alongside emergence angle and profile, in peri-implant disease [16]. Part of the limitation is that the property has no operational definition. Without one, a laboratory cannot report it, design software cannot check it, and studies cannot be pooled.
Digital workflows have removed the practical obstacle to defining it. The geometry of the prosthesis exists as a surface model before fabrication, and the mucosa is captured by an intraoral scan. The question of whether a tool of given shape can reach a given surface without collision is a standard one in manufacturing, where it is called accessibility analysis and is used to plan inspection by coordinate measuring machines [17]. This manuscript transfers that idea to oral hygiene. It defines a cleanability index as the fraction of a critical prosthesis surface reachable by specified cleaning instruments, describes how to compute it, and sets out how it would be validated. A companion proposal in this portfolio concerns passive exchange of saliva around restorations; the present index concerns active instrument access only.
2. Background and Current Scientific Evidence
Plaque, access, and peri-implant disease
The causal role of plaque in mucositis rests on experimental studies in humans [1, 2]. For peri-implantitis the evidence is observational: poor plaque control and lack of regular maintenance are consistent risk indicators [4, 5]. Estimates of prevalence vary with case definition [18]; in a Swedish population sample, moderate or severe peri-implantitis was diagnosed in 14.5% of patients, with higher odds among those with four or more implants [19].
Access for hygiene enters this literature in three forms. The first is direct clinical judgment, as in the Serino and Ström series [10]. A later cross-sectional study classified implants by whether a 0.5-mm interproximal brush could be passed and found that accessibility was associated, to a certain extent, with the peri-implant condition [20]. The second is radiographic contour. Katafuchi and colleagues found peri-implantitis at 31.3% of bone-level implants with emergence angles above 30 degrees and 15.1% of those at or below 30 degrees, and at 37.8% when a wide angle was combined with a convex profile; no such relation was found at tissue-level implants [11]. Yi and colleagues reported similar associations and a higher risk for implants splinted to neighbors on both sides [12]. A systematic review identified only three eligible studies and urged caution about causality [21], and a recent meta-analysis reached conclusions in the same direction on the basis of two articles for each comparison [22]. The third form is implant position: in a cross-sectional analysis of 452 implants, malpositioning in several planes and tilting were strongly associated with peri-implantitis [23]. Malposition is thought to act in part by forcing a restorative contour that obstructs hygiene [14].
Each of these descriptors is a proxy. An emergence angle measured on a periapical radiograph describes one mesial or distal profile in one plane. A brush-passage test describes one instrument at one embrasure. None accounts for the mucosa, the neighboring units, or the path by which an instrument arrives.
Laboratory studies of cleaning
In vitro studies have measured removal of artificial biofilm from implant crowns directly. Tuna and colleagues coated crowns of a conventional and an alternative design with occlusion spray and cleaned them with several interproximal aids. Removal was incomplete with every aid; the cleaned proportion ranged from about three quarters of the surface with a spongy floss to less than one tenth with a microdroplet device, and the alternative design allowed more removal with floss and interdental brush [24]. Sirinirund and colleagues compared concave, straight, and convex crown contours and found that the concave contour allowed the best removal, that floss and spongy floss outperformed an interdental brush and a water flosser, and that no aid removed all of the artificial biofilm [25]. In a clinical experiment on interdental brushes, a waist-shaped brush removed more biofilm than a straight one at the line angles, and neither removed biofilm from buccal sites [26].
These results show that removal depends jointly on the geometry of the restoration and the geometry of the instrument. They were obtained on a few fixed designs. They do not provide a way to predict, for a new design, which surfaces a given aid will miss.
Plaque on full-arch prostheses
The intaglio surface of fixed full-arch prostheses has been studied after retrieval. On 20 maxillary prostheses supported by four implants, plaque covered on average 28% of the fitting surface six months after delivery, three times more on the palatal than on the buccal area, and more in interimplant areas where implants were close together [27]. A three-dimensional method that maps disclosed plaque on a scan of the prosthesis showed far better repeatability than two-dimensional photographic scoring [28]. With such methods, pontic areas that do not contact the mucosa accumulated less plaque than contacting pontics [29], saddle-shaped pontics retained more biofilm than flat or ovate ones, and biofilm area correlated with transmucosal height [30]. Concave intaglio surfaces were associated with more biofilm and bone loss in a cross-sectional study in which only about one fifth of patients were fully satisfied with their ability to clean [31]. Material matters as well: titanium prostheses carried more plaque than zirconia prostheses at all time points in one clinical comparison [32], in keeping with the known influence of surface roughness and surface energy on plaque formation [33, 34].
These studies supply two things that the present proposal needs: a validated way to measure where plaque remains in three dimensions, and evidence that plaque distribution follows geometry.
Self-care evidence
Systematic reviews of self-care around implants find the evidence thin. A review for the European Workshop concluded that the efficacy of powered toothbrushes and of adjunctive chemical agents in managing mucositis remained to be established [35]. A recent review of self-care practices identified twelve studies across five domains of intervention [36]. For natural teeth, interdental brushes have the best support among interdental devices [37, 38]. Recommendations for implant patients are extrapolated from teeth and tailored at the chair to what will fit. An aging implant population adds a further constraint, since prosthetic design for older patients must anticipate declining dexterity and dependence on carers [39].
Accessibility analysis in engineering
In manufacturing, a surface feature is accessible to a tool if the tool can be placed on it from some direction without collision with the part. Algorithms compute, for each surface point, the set of directions from which a probe of given dimensions can approach, and use these sets to plan inspection [17]. The same reasoning underlies tool-path planning for machining and path planning in robotics. The dental problem has features that the manufacturing problem lacks: one boundary, the mucosa, is deformable and must not be injured; some instruments are flexible; and the operator works through a restricted opening. These features determine how the engineering idea has to be adapted and are addressed in Section 5.
Summary
Clinical and laboratory evidence agrees that restricted access is associated with plaque retention and with peri-implant disease, that removal is incomplete even under laboratory conditions, and that geometry governs where plaque remains. Access itself is recorded by clinical judgment or by proxy, and no computed measure was identified.
3. Research Gap
Three gaps follow from this evidence.
First, access for hygiene has no quantitative definition. Studies that use the term record it as a binary judgment [10, 20], and risk assessment tools include prosthesis-related factors as a qualitative item among eight parameters [40]. Emergence angle and profile category are reproducible but describe a single two-dimensional section [11, 21].
Second, laboratory cleaning studies report removal for the designs they test and offer no model that generalizes to other designs [24, 25].
Third, design software for abutments and restorations automates margins, contours, and emergence [41–43]. The patent records retrieved for this manuscript describe design automation; none was found to describe a computed score of cleanability, although the patent search was limited and claims were not read.
PubMed searches combining implant prostheses with cleanability, hygiene access, and accessibility returned clinical and in vitro studies of the kinds reviewed above and no geometric index. A registry search identified a study of access for probing around implants [44], a related but distinct question. The absence of a matching record does not establish that no such index exists, particularly in engineering or commercial software literature that these searches did not cover.
4. Research Objective and Central Hypothesis
The objective is to define, compute, and validate a cleanability index for implant-supported prostheses.
Specific aims are:
- To define the index from the digital geometry of the prosthesis and mucosa and from geometric models of cleaning instruments.
- To implement a computation that returns the index and a map of unreachable surface.
- To determine in vitro whether predicted unreachable surface coincides with residual artificial biofilm after standardized cleaning.
- To determine in a cross-sectional clinical study whether the index is associated with plaque on retrieved prostheses and with peri-implant bleeding.
- To establish the measurement properties of the index: reliability, agreement, and construct validity.
The central hypothesis has two parts. The strong part is nearly a matter of logic and serves as a test of the instrument models: surface that no modeled pose can reach retains biofilm after cleaning, whoever cleans. The weaker and more useful part is empirical: across designs, the index explains more of the variance in residual plaque than emergence angle, profile category, or binary clinical judgment of access. The hypothesis would be refuted if residual biofilm were as common on surface predicted to be reachable as on surface predicted to be unreachable, or if the index added nothing to the simpler descriptors.
No hypothesis about peri-implantitis is tested. Whether a higher index reduces disease is a downstream question that requires longitudinal data.
5. Scientific and Theoretical Framework
Surfaces
Let S be the surface of the prosthesis and abutment exposed to the oral cavity, represented as a triangulated mesh, and let M be the surface of the adjacent mucosa and neighboring teeth or restorations. Not all of S matters equally. The critical surface S_c⊂S is defined as the set of points within a specified distance of the mucosal margin, measured along the surface, together with all points that face the mucosa across a gap smaller than a specified width. The first part captures the transmucosal and juxtamucosal contour; the second captures intaglio and pontic surfaces. The two distances are parameters of the index and are reported with it. Occlusal and free axial surfaces, which a toothbrush reaches without difficulty, are excluded so that they do not dilute the measure.
Instrument models
Each instrument i is represented by three elements. The rigid core is the part that cannot deform and must not intersect the prosthesis: the twisted wire of an interdental brush, the shaft of a single-tuft brush, the nozzle of an irrigator. The working envelope is the region within which the compliant part of the instrument, the bristles or the filament, can exert contact. The approach set is the set of positions and directions from which a hand can present the instrument, limited by the mouth opening, the cheeks, and the tongue. A pose q of the instrument is admissible if the core does not intersect S, does not indent M by more than a tolerance δ, and can be reached from outside the mouth along a continuous path of admissible poses beginning in the approach set. The set of admissible poses is C_i.
Flexible filaments require separate treatment. A floss or tape is threaded through a passage and pulled taut around the restoration. Its admissible configurations are the shortest paths around the obstacle between two held ends, so that in the plane of tension it lies on the convex hull of the cross-section. A spongy floss adds a compliant thickness around that path.
Reach
The reach set of instrument i is the set of surface points that lie within the working envelope W_i(q) for at least one admissible pose:
R_i={ x∈S:∃ q∈C_i with x∈W_i(q) } (1)
Contact alone does not remove biofilm; the working part must move across the surface. A stricter set R_i^* requires that the instrument can be displaced along an admissible path by at least a minimum stroke while the point remains within the envelope. Both are computed; the primary index uses the stricter set.
The index
For a single instrument, the cleanability index is the reachable fraction of the critical surface:
κ_i=(A(R_i^*∩S_c))/(A(S_c)) (2)
where A denotes area. For a kit K of instruments, reach sets are combined:
κ_K=(A(⋃i∈KR_i^*∩S_c))/(A(S_c)) (3)
A weighted form gives more importance to surface near the mucosal margin, where biofilm is in contact with the tissue:
κ_K^w=(∫_S_c^ w(x) 1_(R_K^*)(x) dA)/(∫_S_c^ w(x) dA) (4)
with w a decreasing function of distance from the margin. The unweighted kit index for a reference kit is proposed as the primary reported value. The map of S_c\R_K^*, the unreachable critical surface, is reported alongside it because it shows where a design fails and is more useful to a technician than a number.
Two secondary quantities describe effort. The access count is the smallest number of distinct approach directions needed to realize the kit reach set. The kit size is the smallest number of instruments that achieves a stated fraction of κ_K. A design that reaches the same index with one brush from the buccal side as another reaches with three instruments from four directions is easier to maintain, and the index alone would not show the difference.
Link to residual plaque
Let P be the fraction of the critical surface that carries biofilm after a cleaning episode, starting from full coverage. Unreached surface stays covered. On reached surface a fraction η is removed, where η depends on instrument, time, force, and operator. Then
P=(1−κ_K)+κ_K(1−η)=1−η κ_K (5)
Equation (5) is the measurement model. It predicts that residual plaque falls linearly with the index, with a slope equal to the removal efficiency, and that the residual never falls below 1−κ_K. In vitro data showing that even the best aid left about one quarter of the artificial biofilm on single crowns [24] indicate that η is well below one or that reach is incomplete, and the model allows the two to be separated.
6. Methodology
Computation
Surfaces are imported as triangulated meshes and remeshed to a uniform edge length chosen by a convergence study, in which the index is recomputed at successively finer resolutions until it changes by less than a stated tolerance. The prosthesis and mucosa are registered in the coordinate frame of the intraoral scan.
For rigid straight instruments, the admissible set is searched by sampling. Candidate axes are generated over entry points on a bounding surface around the site and over directions within the approach set. Each candidate is tested for collision of the core with the prosthesis and for indentation of the mucosa. For each admissible axis, surface points within the working envelope are marked as reached, and the strict criterion is tested by translating and rotating the axis within the admissible set. Curved or angled instruments are handled by a sampling-based path planner that tests whether the pose can be reached along a continuous collision-free path. For filaments, passages that connect two regions accessible to the fingers or to a threader are identified, the taut path is computed as a geodesic on the convex hull of successive cross-sections, and the compliant thickness is added for spongy floss.
The approach set is derived from a simplified envelope of the oral opening defined by interincisal opening, commissure position, and the retracted cheek. Its parameters are not fixed here. They would be taken from anthropometric sources and varied in sensitivity analysis, and the dependence of the index on them would be reported.
Instrument library
Instrument models require measured dimensions. Table 1 lists the parameters for each class. Values are obtained by bench measurement of commercial instruments: core diameter by micrometer, envelope diameter by optical profilometry of the unloaded brush, usable length from the handle geometry, and bristle compliance by a transverse force–deflection test that sets the envelope within which a stated contact force is maintained. No parameter values are asserted in this manuscript.
| Instrument class | Rigid core | Working envelope | Approach constraint | Parameters to be measured |
|---|---|---|---|---|
| Interdental brush, straight | Wire | Cylinder about the wire | Straight insertion from buccal or lingual | Wire diameter; envelope diameter under stated contact force; usable length |
| Interdental brush, angled or waist-shaped | Wire with bend | Profiled envelope | Insertion after rotation of handle | As above, with bend angle and profile |
| Single-tuft brush | Shaft and ferrule | Cone of bristles | Free within oral envelope | Tuft diameter and length; shaft angle |
| Toothbrush | Head | Field of bristles | Free within oral envelope | Head dimensions; bristle length |
| Floss or tape | None | Line on convex hull | Requires a through-passage or contact that can be passed | Thickness |
| Spongy floss | Stiffened end as threader | Convex-hull path thickened by compressed sponge | Requires a through-passage | Compressed thickness under stated tension |
| Oral irrigator | Nozzle | Line of sight from nozzle within stand-off range | Free within oral envelope | Nozzle diameter; effective range |
Irrigation is included for completeness as a line-of-sight model. Removal by water jets depends on fluid mechanics that a geometric model does not represent [45], and the irrigator is excluded from the reference kit.
Stage 1: in vitro validation
Prostheses are designed to span a wide range of predicted index values by varying emergence angle, profile, embrasure form, pontic form, and the height of the intaglio above the mucosa. Single crowns, three-unit splinted restorations, and segments of full-arch prostheses are included. Each is fabricated and mounted on a model with a silicone mucosal analog of defined stiffness. Surfaces are coated with an artificial biofilm analog of the kind used in previous in vitro studies [24, 25].
Cleaning is performed in two ways. In the mechanized condition, a positioning stage moves each instrument through the admissible poses identified by the computation, which tests the instrument models without operator variability. In the manual condition, trained operators blinded to the predicted maps clean each specimen with a specified instrument for a fixed time. Each specimen is then scanned in color and residual coating is mapped on the surface model, following the three-dimensional approach that showed high repeatability for plaque on full-arch prostheses [28].
The primary outcome is point-by-point agreement between the predicted unreachable surface and the observed residual coating, summarized by sensitivity, specificity, and an overlap coefficient. The secondary outcome is the relation between the residual fraction and the index across specimens, to which equation (5) is fitted.
Stage 2: clinical cross-sectional study
Participants are adults with screw-retained implant prostheses that are removed as part of scheduled maintenance. Before removal, modified plaque and bleeding indices are recorded at each implant, and an intraoral scan is taken. After removal, plaque on the prosthesis is disclosed and scanned, and the mucosa is scanned without the prosthesis. The index is computed from the scans by an analyst blinded to clinical findings. Plaque coverage of the critical surface is measured from the color scan, and conventional plaque scores are recorded for comparison [46, 47].
Covariates include self-reported hygiene aids and frequency, time since the last professional cleaning, prosthesis material, manual dexterity, and history of periodontitis. Emergence angle and profile are measured from radiographs as in earlier studies [11, 12], and a clinician records a binary judgment of access [10] before the index is known.
The primary analysis relates plaque coverage to the index for the kit that each participant reports using. A second analysis uses the reference kit and asks how much of the plaque distribution is explained by geometry alone. Bleeding on probing is a secondary outcome analyzed at implant level with adjustment for clustering within patients.
Measurement properties
Measurement properties are assessed against published standards for health measurement instruments [48]. Reliability of the full procedure, from scan to index, is estimated from repeated scans by different operators and expressed as an intraclass correlation coefficient whose form is selected and reported according to guidance [49]. Agreement between repeated measurements is displayed by the method of Bland and Altman [50]. Construct validity is tested through prespecified hypotheses: the index is lower for restorations with emergence angles above 30 degrees and convex profiles; lower for saddle than for ovate pontics; lower at implants splinted on both sides; and lower at sites that a clinician judges inaccessible. Criterion validity is addressed by Stage 1, in which residual coating after mechanized cleaning is the reference.
| Property | Stage | Reference or comparison | Analysis | Proposed criterion |
|---|---|---|---|---|
| Criterion validity of reach maps | 1, mechanized | Residual coating after exhaustive mechanized cleaning | Sensitivity and specificity of predicted unreachable surface | Both above a prespecified threshold |
| Predictive validity | 1, manual | Residual fraction after timed manual cleaning | Fit of equation (5); comparison with emergence angle and profile | Index explains more variance than comparators |
| Reliability | 1 and 2 | Repeated scans and analysts | Intraclass correlation coefficient | Lower confidence bound above a prespecified value |
| Agreement | 1 and 2 | Repeated measurement | Limits of agreement | Limits narrower than a stated fraction of the index range |
| Construct validity | 2 | Emergence angle, profile, pontic form, clinical judgment | Prespecified directional hypotheses | At least three quarters confirmed |
| Clinical association | 2 | Plaque coverage; bleeding on probing | Mixed-effects regression | Estimate with confidence interval; no threshold |
Sample size
For Stage 1, the number of specimens is set by the precision of the fitted slope in equation (5) and by the reliability estimate. Thirty designs, each cleaned in replicate with each instrument by three operators, is the planning figure, to be confirmed by simulation once pilot variances are known. For Stage 2, detecting a correlation of 0.5 between index and plaque coverage with 80% power at a two-sided significance level of 0.05 requires about 29 independent prostheses by the Fisher transformation. Because the true correlation is unknown and implants cluster within patients, the target would be set higher and fixed after the pilot. These figures are planning arithmetic and are not based on data.
Statistical analysis
Analyses are prespecified. Point-level agreement in Stage 1 is computed per specimen and summarized across specimens with confidence intervals from bootstrap resampling of specimens. Regression of residual fraction on the index includes instrument and operator as factors. In Stage 2, mixed-effects models with random intercepts for patient account for clustering. The added value of the index over emergence angle, profile, and clinical judgment is assessed by comparing nested models. Sensitivity analyses vary the definition of the critical surface, the mucosal tolerance, and the approach set.
Reproducibility
Code, instrument models, test geometries, and the Stage 1 designs would be released openly so that other groups can compute the index on their own designs and challenge the instrument models.
Ethics
Stage 1 involves no human participants. Stage 2 requires approval by an institutional review board and written informed consent. Procedures are those of routine maintenance with the addition of scans and disclosure; removal of prostheses solely for research is not proposed. No approval has been sought, because no study has been initiated.
7. Analytical Development
Closed-form cases show how the index behaves and provide checks for the computation. The dimensions used are chosen for illustration and are not measurements.
Filament in a concavity
A taut filament wrapped around a restoration lies, in the plane of tension, on the convex hull of the cross-section. It touches the surface where the surface coincides with the hull and bridges every concavity. If the perimeter of the cross-section within the critical surface is p and a length a of it lies in concavities, the filament-reachable fraction in that section is 1−a/p. A section of perimeter 30 mm with 6 mm of concave contour has a reachable fraction of 0.80. The mucosa-facing surface of a saddle pontic is concave in the buccolingual plane and is bridged entirely.
A spongy floss of compressed thickness t follows the same path and fills concavities shallower than t. For a concavity with chord c and maximum depth s, approximated as a parabola, the depth at distance x from the center is s (1−4x^2/c^2), and the fraction of the chord over which the depth is within t is
f=1−√(1−t/s), t<s (6)
and f=1 for t≥s. For a depth of 2 mm and a compressed thickness of 1 mm, f=0.29: the sponge reaches the shoulders of the concavity and misses most of its floor. For a depth of 1.5 mm, f=0.42. When the depth does not exceed the thickness, the whole concavity is reached. The model thus predicts a threshold effect: concavities are cleanable by spongy floss when shallower than its compressed thickness and largely uncleanable when deeper.
This result concerns concavity in the plane in which the filament is under tension. It should not be confused with a concave emergence profile seen in a vertical section, which enlarges the embrasure and was associated with better removal in vitro [25]. A crown with a concave vertical profile usually has convex horizontal sections, which a filament contacts fully. The distinction shows why a single profile category cannot stand in for a three-dimensional computation.
Brush beneath a full-arch prosthesis
Consider the intaglio surface of a full-arch prosthesis between two implants as a strip of buccolingual width W at height h above the mucosa. A straight interdental brush has core diameter d_c, envelope diameter d_b, and usable length L. Three regimes follow from the height. If h+δ<d_c, the core cannot enter and nothing is reached. If d_c≤h+δ≤d_b, the bristles bear on the intaglio and on the mucosa together. If h>d_b, the brush enters freely and reaches the intaglio only when lifted against it, which requires that the handle can be angled within the approach set.
Within the second and third regimes, the reached fraction of the strip width depends on whether the brush can be inserted from one side or from both:
κ=min(1,mL/W), m∈{1,2} (7)
Table 3 evaluates equation (7) for a usable length of 10 mm.
| Strip width W (mm) | Insertion from one side | Insertion from both sides |
|---|---|---|
| 8 | 1.00 | 1.00 |
| 12 | 0.83 | 1.00 |
| 16 | 0.63 | 1.00 |
| 24 | 0.42 | 0.83 |
The calculation makes a simple point. Where the palatal or lingual side cannot be approached, the unreached band lies on that side, and it widens in proportion to the width of the prosthesis. The clinical finding that plaque on maxillary full-arch prostheses was about three times more extensive on the palatal than on the buccal part of the fitting surface [27] is in the direction this model predicts. The agreement is qualitative. Other explanations, including differences in salivary clearance and in how patients direct their efforts, are equally compatible with that observation.
The mucosal tolerance
The tolerance δ is the depth to which an instrument may displace the mucosa. With δ=0 a contacting pontic has no reachable intaglio at all. With a tolerance of a fraction of a millimeter, a thin floss may pass. The index is therefore sensitive to δ where gaps are near the core diameter, and this sensitivity is a property of the clinical situation, in which patients do force instruments beneath pontics. The proposal is to report the index at two tolerances, zero and a value derived from the indentation produced by a stated force on the mucosal analog, and to treat a large difference between them as a flag that cleaning depends on tissue displacement.
What the geometric model omits
The model treats removal on reached surface through a single efficiency. It does not represent adhesion strength of biofilm, surface roughness, or material [32–34], nor the skill, vision, and persistence of the person cleaning. It predicts where cleaning is possible and is silent on whether it happens.
8. Discussion
The proposal converts a clinical impression into a quantity that can be computed before a prosthesis is made. Its justification is the convergence of three lines of evidence: clinical association between restricted access and peri-implantitis [10, 11, 23], laboratory demonstration that removal depends on contour and instrument [24, 25], and mapping studies showing that plaque on retrieved prostheses follows geometry [27, 29, 30]. The index attempts to supply the variable that these studies approach from different sides.
The index should be judged first as a measurement. A measure of access is useful only if it is reproducible and if it predicts something that simpler measures do not. Emergence angle is a strong competitor. It is easy to obtain, and its association with peri-implantitis has been replicated, although on few studies and only at bone-level implants [21, 22]. The reported absence of an association at tissue-level implants [11] is a useful test case. If the angle acts through access, an index that accounts for the position of the contour relative to the mucosa should explain why the same angle matters in one configuration and not in the other. If it cannot, either the index is deficient or the angle acts through another pathway.
The measurement model in equation (5) also has a consequence for how in vitro cleaning studies are read. A reported cleaned fraction combines reach and efficiency. Two aids with the same cleaned fraction may differ entirely in what they miss: one reaching everywhere and removing partially, the other removing completely where it reaches and missing a concavity altogether. For a patient who cleans daily, partial removal on reached surface is compensated by repetition, whereas unreached surface is never cleaned. The distinction matters for disease and is invisible in a single percentage.
Several objections deserve attention. One is that patients do not clean as a planner would. A design with a high index for a kit of four instruments used from four directions may be worse in practice than a design with a moderate index for one brush. The access count and kit size are intended to capture this, and the clinical stage analyzes the instruments that participants report using. A second objection is that the instrument models may be wrong in ways that matter, particularly for bristles, which splay and bend. The mechanized arm of Stage 1 exists to test the models in isolation from operator behavior, and the open release of models invites correction. A third is that plaque is not disease. An index validated against plaque has not been shown to predict mucositis or bone loss, and host factors may dominate. The present proposal stops at plaque and bleeding and makes no claim beyond them.
A further question is whether a single number is the right output. A designer needs to know where a design fails and what change would repair it; the map of unreachable surface answers that directly. A summary index is needed for research, for comparison of designs, and for any future threshold. Both are provided, and the map is likely to be the more useful of the two in the laboratory.
9. Clinical Significance
If validated, the index would have three uses. At the design stage, it would give the technician and prosthodontist a check on cleanability comparable to existing checks on thickness and occlusal clearance, applied before fabrication. At delivery, the reach maps would support specific hygiene instruction: which aid, at which site, from which side. For a patient with reduced dexterity or one who depends on a carer, the kit could be restricted to instruments that are realistic and the design adjusted to suit them. In research, the index would provide a covariate for studies of prosthesis design and peri-implant disease that is defined identically across centers.
None of these uses is established. The index does not replace clinical judgment, and no threshold separating acceptable from unacceptable designs can be proposed without longitudinal data.
10. Limitations
The index has not been computed on any clinical geometry, and the closed-form cases are idealizations. Instrument parameters are unmeasured. The mucosa is treated as a boundary with a displacement tolerance, whereas real tissue varies in stiffness and sensitivity. The approach set is a simplification of a hand working in a mouth. Subgingival surfaces within the peri-implant sulcus are not addressed, although they are where disease occurs. The scans on which the index depends have limited accuracy in narrow gaps, which are exactly the regions that determine the result. The clinical stage is cross-sectional and can show association only. All sources were assessed from abstracts or index records, and details such as the composition of the artificial biofilm analogs and the criteria used for clinical judgments of access would have to be taken from the full texts.
11. Future Validation
Execution of Stages 1 and 2 is the immediate requirement. Beyond them, a longitudinal cohort would relate the index at delivery to incident mucositis and bone loss, with adjustment for patient-level risk. A design-intervention study would compare prostheses designed with and without feedback from the index, with plaque as the outcome. Extension to removable and overdenture attachments, to orthodontic appliances, and to natural-tooth restorations is possible in principle. Agreement on a reference kit and on the definition of the critical surface would require consensus among clinicians and manufacturers.
12. Conclusion
Access for oral hygiene is repeatedly associated with peri-implant disease and has not been defined in measurable terms. A cleanability index based on the reachable fraction of a critical prosthesis surface, computed from digital geometry and from geometric models of cleaning instruments, would provide such a definition. Simple cases show that the index distinguishes surfaces a filament bridges from those it touches and that it predicts where brushes fail beneath wide prostheses. Whether it predicts residual plaque better than existing descriptors is an empirical question, and the protocol described here is designed to answer it. No data have been collected.
Cite this paper
Jain S, Gupta A. A Quantitative Cleanability Index for Implant-Supported Prostheses: Geometric Definition, Computational Method, and a Protocol for Validation Against Residual Plaque. Working paper. Center for Implant Dentistry; 2026. Available from: https://bayareaimplantdentistry.com/clinician-research-and-publications/quantitative-cleanability-index
References
- Pontoriero R, Tonelli MP, Carnevale G, Mombelli A, Nyman SR, Lang NP. Experimentally induced peri-implant mucositis. A clinical study in humans. Clin Oral Implants Res. 1994;5(4):254-9. doi:10.1034/j.1600-0501.1994.050409.x (opens in a new tab). PMID: 7640340 (opens in a new tab).
- Salvi GE, Aglietta M, Eick S, Sculean A, Lang NP, Ramseier CA. Reversibility of experimental peri-implant mucositis compared with experimental gingivitis in humans. Clin Oral Implants Res. 2012;23(2):182-190. doi:10.1111/j.1600-0501.2011.02220.x (opens in a new tab). PMID: 21806683 (opens in a new tab).
- Heitz-Mayfield LJA, Salvi GE. Peri-implant mucositis. J Periodontol. 2018;89 Suppl 1:S257-S266. doi:10.1002/JPER.16-0488 (opens in a new tab). PMID: 29926954 (opens in a new tab).
- Schwarz F, Derks J, Monje A, Wang HL. Peri-implantitis. J Periodontol. 2018;89 Suppl 1:S267-S290. doi:10.1002/JPER.16-0350 (opens in a new tab). PMID: 29926957 (opens in a new tab).
- Berglundh T, Armitage G, Araujo MG, Avila-Ortiz G, Blanco J, Camargo PM, et al. Peri-implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions. J Clin Periodontol. 2018;45 Suppl 20:S286-S291. doi:10.1111/jcpe.12957 (opens in a new tab). PMID: 29926491 (opens in a new tab).
- Jepsen S, Berglundh T, Genco R, Aass AM, Demirel K, Derks J, et al. Primary prevention of peri-implantitis: managing peri-implant mucositis. J Clin Periodontol. 2015;42 Suppl 16:S152-7. doi:10.1111/jcpe.12369 (opens in a new tab). PMID: 25626479 (opens in a new tab).
- Herrera D, Berglundh T, Schwarz F, Chapple I, Jepsen S, Sculean A, et al. Prevention and treatment of peri-implant diseases-The EFP S3 level clinical practice guideline. J Clin Periodontol. 2023;50 Suppl 26:4-76. doi:10.1111/jcpe.13823 (opens in a new tab). PMID: 37271498 (opens in a new tab).
- Costa FO, Takenaka-Martinez S, Cota LO, Ferreira SD, Silva GL, Costa JE. Peri-implant disease in subjects with and without preventive maintenance: a 5-year follow-up. J Clin Periodontol. 2012;39(2):173-81. doi:10.1111/j.1600-051X.2011.01819.x (opens in a new tab). PMID: 22111654 (opens in a new tab).
- Monje A, Aranda L, Diaz KT, Alarcón MA, Bagramian RA, Wang HL, et al. Impact of Maintenance Therapy for the Prevention of Peri-implant Diseases: A Systematic Review and Meta-analysis. J Dent Res. 2016;95(4):372-9. doi:10.1177/0022034515622432 (opens in a new tab). PMID: 26701350 (opens in a new tab).
- Serino G, Ström C. Peri-implantitis in partially edentulous patients: association with inadequate plaque control. Clin Oral Implants Res. 2009;20(2):169-74. doi:10.1111/j.1600-0501.2008.01627.x (opens in a new tab). PMID: 19077152 (opens in a new tab).
- Katafuchi M, Weinstein BF, Leroux BG, Chen YW, Daubert DM. Restoration contour is a risk indicator for peri-implantitis: A cross-sectional radiographic analysis. J Clin Periodontol. 2018;45(2):225-232. doi:10.1111/jcpe.12829 (opens in a new tab). PMID: 28985447 (opens in a new tab).
- Yi Y, Koo KT, Schwarz F, Ben Amara H, Heo SJ. Association of prosthetic features and peri-implantitis: A cross-sectional study. J Clin Periodontol. 2020;47(3):392-403. doi:10.1111/jcpe.13251 (opens in a new tab). PMID: 31912511 (opens in a new tab).
- Monje A, Insua A, Wang HL. Understanding Peri-Implantitis as a Plaque-Associated and Site-Specific Entity: On the Local Predisposing Factors. J Clin Med. 2019;8(2). doi:10.3390/jcm8020279 (opens in a new tab). PMID: 30823574 (opens in a new tab).
- Hamilton A, Putra A, Nakapaksin P, Kamolroongwarakul P, Gallucci GO. Implant prosthodontic design as a predisposing or precipitating factor for peri-implant disease: A review. Clin Implant Dent Relat Res. 2023;25(4):710-722. doi:10.1111/cid.13183 (opens in a new tab). PMID: 36691784 (opens in a new tab).
- Berglundh T, Mombelli A, Schwarz F, Derks J. Etiology, pathogenesis and treatment of peri-implantitis: A European perspective. Periodontol 2000. 2024. doi:10.1111/prd.12549 (opens in a new tab). PMID: 38305506 (opens in a new tab).
- Mattheos N, Janda M, Acharya A, Pekarski S, Larsson C. Impact of design elements of the implant supracrestal complex (ISC) on the risk of peri-implant mucositis and peri-implantitis: A critical review. Clin Oral Implants Res. 2021;32 Suppl 21:181-202. doi:10.1111/clr.13823 (opens in a new tab). PMID: 34642979 (opens in a new tab).
- Spitz SN, Spyridi AJ, Requicha AG. Accessibility analysis for planning of dimensional inspection with coordinate measuring machines. IEEE Trans. Robot. Automat.. 1999;15(4):714-727. doi:10.1109/70.782025 (opens in a new tab).
- Derks J, Tomasi C. Peri-implant health and disease. A systematic review of current epidemiology. J Clin Periodontol. 2015;42 Suppl 16:S158-71. doi:10.1111/jcpe.12334 (opens in a new tab). PMID: 25495683 (opens in a new tab).
- Derks J, Schaller D, Håkansson J, Wennström JL, Tomasi C, Berglundh T. Effectiveness of Implant Therapy Analyzed in a Swedish Population: Prevalence of Peri-implantitis. J Dent Res. 2016;95(1):43-9. doi:10.1177/0022034515608832 (opens in a new tab). PMID: 26701919 (opens in a new tab).
- Pons R, Nart J, Valles C, Salvi GE, Monje A. Self-administered proximal implant-supported hygiene measures and the association to peri-implant conditions. J Periodontol. 2021;92(3):389-399. doi:10.1002/JPER.20-0193 (opens in a new tab). PMID: 32761897 (opens in a new tab).
- Soulami S, Slot DE, van der Weijden F. Implant-abutment emergence angle and profile in relation to peri-implantitis: A systematic review. Clin Exp Dent Res. 2022;8(4):795-806. doi:10.1002/cre2.594 (opens in a new tab). PMID: 35713938 (opens in a new tab).
- Lin GH, Lee E, Barootchi S, Rosen PS, Curtis D, Kan J, et al. The influence of prosthetic designs on peri-implant bone loss: An AO/AAP systematic review and meta-analysis. J Periodontol. 2025;96(6):634-651. doi:10.1002/JPER.24-0144 (opens in a new tab). PMID: 40489290 (opens in a new tab).
- Monje A, Pons R, Soldini MC, Rosen PS, Nart J, Buser D. Role of Local Factors on the Occurrence of Peri-Implantitis: A Cross-Sectional Analysis. Clin Oral Implants Res. 2025;36(12):1614-1625. doi:10.1111/clr.70036 (opens in a new tab). PMID: 40916656 (opens in a new tab).
- Tuna T, Kuhlmann L, Bishti S, Sirazitdinova E, Deserno T, Wolfart S. Removal of simulated biofilm at different implant crown designs with interproximal oral hygiene aids: An in vitro study. Clin Oral Implants Res. 2019;30(7):627-636. doi:10.1111/clr.13448 (opens in a new tab). PMID: 31033028 (opens in a new tab).
- Sirinirund B, Siqueira R, Li J, Mendonça G, Zalucha J, Wang HL. Effects of crown contour on artificial biofilm removal efficacy with interdental cleaning aids: An in vitro study. Clin Oral Implants Res. 2023;34(8):783-792. doi:10.1111/clr.14105 (opens in a new tab). PMID: 37269176 (opens in a new tab).
- Chongcharoen N, Lulic M, Lang NP. Effectiveness of different interdental brushes on cleaning the interproximal surfaces of teeth and implants: a randomized controlled, double-blind cross-over study. Clin Oral Implants Res. 2012;23(5):635-40. doi:10.1111/j.1600-0501.2011.02387.x (opens in a new tab). PMID: 22146002 (opens in a new tab).
- Abi Nader S, Eimar H, Momani M, Shang K, Daniel NG, Tamimi F. Plaque Accumulation Beneath Maxillary All-on-4™ Implant-Supported Prostheses. Clin Implant Dent Relat Res. 2015;17(5):932-7. doi:10.1111/cid.12199 (opens in a new tab). PMID: 24461161 (opens in a new tab).
- Jia PY, Liu Q, Yang JW, Jiang XT, Afrashtehfar KI. Three-dimensional macroscopical quantification of plaque accumulation on implant-supported fixed complete dental prostheses surfaces: A methodological pilot study. J Prosthet Dent. 2022;128(3):467.e1-467.e8. doi:10.1016/j.prosdent.2021.11.008 (opens in a new tab). PMID: 35940952 (opens in a new tab).
- Gong Z, Lin Y, Di P. Plaque accumulation on the fitting surface of full-arch implant-supported fixed prostheses with contact or noncontact pontics: A split mouth randomized controlled trial. J Esthet Restor Dent. 2023;35(7):1077-1084. doi:10.1111/jerd.13062 (opens in a new tab). PMID: 37171039 (opens in a new tab).
- Jia P, Yang J. Biofilm accumulation on complete arch subperiosteal implant-supported prostheses: Effect of pontic type. J Prosthet Dent. 2024;131(3):478.e1-478.e6. doi:10.1016/j.prosdent.2023.12.015 (opens in a new tab). PMID: 38233257 (opens in a new tab).
- Able FB, Campanha NH, Younes IA, Sartori IAM. Evaluation of the intaglio surface shape of implant-supported complete-arch maxillary prostheses and its association with biological complications: An analytical cross-sectional study. J Prosthet Dent. 2022;128(2):174-180. doi:10.1016/j.prosdent.2020.12.028 (opens in a new tab). PMID: 33563467 (opens in a new tab).
- Curiel-Aguilera FP, Griffiths GR, Rossmann JA, Gonzalez JA. Titanium versus zirconia complete arch implant-supported fixed prostheses: A comparison of plaque accumulation. J Prosthet Dent. 2023;130(3):369-375. doi:10.1016/j.prosdent.2021.10.012 (opens in a new tab). PMID: 34879988 (opens in a new tab).
- Quirynen M, Bollen CM. The influence of surface roughness and surface-free energy on supra- and subgingival plaque formation in man. A review of the literature. J Clin Periodontol. 1995;22(1):1-14. doi:10.1111/j.1600-051x.1995.tb01765.x (opens in a new tab). PMID: 7706534 (opens in a new tab).
- Teughels W, Van Assche N, Sliepen I, Quirynen M. Effect of material characteristics and/or surface topography on biofilm development. Clin Oral Implants Res. 2006;17 Suppl 2:68-81. doi:10.1111/j.1600-0501.2006.01353.x (opens in a new tab). PMID: 16968383 (opens in a new tab).
- Salvi GE, Ramseier CA. Efficacy of patient-administered mechanical and/or chemical plaque control protocols in the management of peri-implant mucositis. A systematic review. J Clin Periodontol. 2015;42 Suppl 16:S187-201. doi:10.1111/jcpe.12321 (opens in a new tab). PMID: 25495416 (opens in a new tab).
- Eagle IT, Theis-Mahon N, Arnett MC. Best oral self-care practices for peri-implant conditions and diseases: a systematic review. Front Oral Health. 2025;6:1657025. doi:10.3389/froh.2025.1657025 (opens in a new tab). PMID: 41487703 (opens in a new tab).
- Slot DE, Dörfer CE, Van der Weijden GA. The efficacy of interdental brushes on plaque and parameters of periodontal inflammation: a systematic review. Int J Dent Hyg. 2008;6(4):253-64. doi:10.1111/j.1601-5037.2008.00330.x (opens in a new tab). PMID: 19138177 (opens in a new tab).
- Van der Weijden GAF, van Loveren C. Mechanical plaque removal in step-1 of care. Periodontol 2000. 2023. doi:10.1111/prd.12541 (opens in a new tab). PMID: 38148481 (opens in a new tab).
- Yeh YT, Chiou LL, Yoon D, Lieuw K, Yang B, Maniewicz S, et al. Implant therapy in older patients: Prosthetic considerations and biomechanical implications for frailty/aging. Periodontol 2000. 2026. doi:10.1111/prd.70058 (opens in a new tab). PMID: 42175573 (opens in a new tab).
- Heitz-Mayfield LJA, Heitz F, Lang NP. Implant Disease Risk Assessment IDRA-a tool for preventing peri-implant disease. Clin Oral Implants Res. 2020;31(4):397-403. doi:10.1111/clr.13585 (opens in a new tab). PMID: 32003037 (opens in a new tab).
- Astra Tech Inc, applicant. Method and system for designing custom restorations for dental implants. European patent application EP 1 973 491 A1. Priority 2006 Jan 5; published 2008 Oct 1 (related CN 101400317 A).
- Dental-abutment design system and method. United States patent application publication US 2020/0138551 A1. Priority 2018 Nov 6.
- Dentsply Sirona Inc, applicant. Method, system and devices for instant automated design of a customized dental object. United States patent application publication US 2022/0296344 A1. Priority 2019 Sep 5; published 2022 Sep 22 (granted as US 12,465,469 B2; related WO 2021/046147 A1).
- ClinicalTrials.gov. Identifier NCT07557459. Access to probing and peri-implant diseases. Bethesda (MD): National Library of Medicine (US).
- Rmaile A, Carugo D, Capretto L, Aspiras M, De Jager M, Ward M, et al. Removal of interproximal dental biofilms by high-velocity water microdrops. J Dent Res. 2014;93(1):68-73. doi:10.1177/0022034513510945 (opens in a new tab). PMID: 24170371 (opens in a new tab).
- QUIGLEY GA, HEIN JW. Comparative cleansing efficiency of manual and power brushing. J Am Dent Assoc. 1962;65:26-9. doi:10.14219/jada.archive.1962.0184 (opens in a new tab). PMID: 14489483 (opens in a new tab).
- Turesky S, Gilmore ND, Glickman I. Reduced plaque formation by the chloromethyl analogue of victamine C. J Periodontol. 1970;41(1):41-3. doi:10.1902/jop.1970.41.41.41 (opens in a new tab). PMID: 5264376 (opens in a new tab).
- Mokkink LB, Terwee CB, Patrick DL, Alonso J, Stratford PW, Knol DL, et al. The COSMIN checklist for assessing the methodological quality of studies on measurement properties of health status measurement instruments: an international Delphi study. Qual Life Res. 2010;19(4):539-49. doi:10.1007/s11136-010-9606-8 (opens in a new tab). PMID: 20169472 (opens in a new tab).
- Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15(2):155-63. doi:10.1016/j.jcm.2016.02.012 (opens in a new tab). PMID: 27330520 (opens in a new tab).
- Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. PMID: 2868172 (opens in a new tab).
- U.S. Food and Drug Administration. Product classification: product code LLZ (medical image management and processing system), 21 CFR 892.2050; 510(k) summaries K152078 (3Shape Implant Studio) and K103723 (SICAT Implant). Silver Spring (MD): FDA.
- U.S. Food and Drug Administration. Product classifications: endosseous dental implant, 21 CFR 872.3640 (product code DZE); endosseous dental implant abutment, 21 CFR 872.3630 (product code NHA). Silver Spring (MD): FDA.