Salivary Microfluidics Around Implant Restorations: A Computational and Experimental Methodology for Relating Prosthetic Geometry to Fluid Exchange, Microbial Retention, and Local Disease Risk
- Paper type
- Computational modeling methodology with in vitro validation protocol
- Status
- Original Work · Working Paper
Contents
Abstract
Background. The contour of an implant restoration is associated with peri-implantitis. In a cross-sectional radiographic study, bone-level implants with an emergence angle above 30° had a peri-implantitis prevalence of 31.3% against 15.1% for smaller angles, and a convex profile compounded the effect. The explanation offered is plaque retention, but the physical process by which contour produces retention has not been modeled. Computational fluid dynamics has related salivary flow around orthodontic brackets to bacterial retention. No comparable study of implant restorations was identified.
Objective. To define the fluid-mechanical regime of saliva around implant restorations, to propose geometry-dependent measures of fluid exchange, and to specify a computational and experimental program that tests whether those measures predict where biofilm is retained.
Framework. Scale analysis places the resting salivary film in the creeping-flow regime, with wall shear stresses far below those that detach biofilm. Under these conditions geometry acts through mass transport: sheltered regions exchange solutes and cells with the bulk by diffusion, on time scales that grow with the square of niche depth. A retention index is defined from the simulated fluid age and exchange rate over the restoration surface.
Proposed methods. A parametric family of emergence profiles, embrasures, and pontic clearances is simulated under resting and functional flow with shear-thinning saliva rheology. Three-dimensional printed replicas are tested in flow cells with a multispecies biofilm, and measured biovolume maps are compared with predicted exchange maps. A clinical study relates the index, computed from design files, to site-level plaque and inflammation.
Central hypothesis. At equal emergence angle, profile convexity and embrasure geometry independently determine the volume of poorly exchanged fluid, and that volume co-localizes with biofilm retention.
Conclusions. The methodology is proposed and untested. The scale analysis indicates that models based on shear removal by resting saliva would be misdirected. No results are reported.
Keywords: implant-supported prosthesis; emergence profile; saliva; computational fluid dynamics; biofilm; residence time; peri-implantitis; prosthesis design
1. Introduction
An implant is round and narrow; the tooth it replaces is neither. The restoration must therefore flare from the implant platform to the dimensions of a crown within a few millimeters of height, and the shape of that transition, the emergence profile, is under the control of the clinician and the laboratory. For most of the history of implant dentistry it was designed for esthetics and soft-tissue support [1, 2]. Evidence now suggests that it also affects disease risk.
Katafuchi and colleagues measured the emergence angle and profile of 168 implant restorations in 83 patients on radiographs. Among bone-level implants, peri-implantitis was present at 31.3% of those with an emergence angle above 30° and at 15.1% of those at or below it; when a wide angle was combined with a convex profile, prevalence was 37.8%; tissue-level implants showed no such association [3]. A second cross-sectional study related prosthetic features to peri-implantitis [4]. A critical review concluded that over-contoured designs were consistently associated with disease [5], and subsequent systematic reviews, including a joint AO/AAP review, found emergence angles below 30° and concave or straight profiles linked to lower risk [6, 7]. Prosthetic design is discussed as a predisposing or precipitating factor for peri-implant disease [8–10].
The mechanism is assumed to be biofilm. A wide, convex transition creates a sheltered zone between the restoration and the mucosa that hygiene devices do not reach and in which plaque matures. That account has two components that are usually run together. One is active: whether a brush or floss can physically reach a surface. The other is passive: whether saliva, which bathes every oral surface and continuously removes bacteria, nutrients, and metabolic products, exchanges effectively with the fluid in the sheltered zone. The first is a question of instrument geometry. The second is a question of fluid mechanics, and it is the subject of this manuscript.
Fluid mechanics has been brought to a similar problem in orthodontics. Simulations of salivary flow over teeth with brackets found that the appliance disturbs the flow field and creates local recirculation gingival to the bracket, where plaque and inflammation are observed clinically [11], and a study using patient-derived models related simulated flow patterns to bacterial retention at specific sites around different bracket designs [12]. A PubMed search for computational fluid dynamics or wall shear stress in connection with implant restorations, emergence profiles, or crowns and biofilm or saliva returned three records, of which only the orthodontic study was relevant (Appendix A).
The proposal is to do for the implant restoration what has been done for the bracket, with one difference in emphasis that follows from a scale analysis presented below: in the thin, slow salivary film, geometry acts mainly by limiting exchange, and the appropriate quantities are residence time and transport rate.
2. Background and Current Scientific Evidence
Contour and peri-implant disease
The epidemiological signal is consistent in direction and modest in quality. The index study was cross-sectional, measured angles on two-dimensional radiographs in the mesial and distal planes only, and included mainly platform-matched bone-level implants [3]. Reviews note that the supracrestal complex comprises several design elements (emergence angle, profile, the height and shape of the transmucosal component, margin position) that are correlated in practice and difficult to separate in observational data [5, 7]. An emergence angle is a single number describing a three-dimensional surface, and two restorations with the same angle can differ in convexity, in embrasure form, and in how the contour varies around the circumference. Inadequate access for hygiene was strongly associated with peri-implantitis in an earlier study [13], and access and contour are themselves related. None of the clinical studies measured fluid exchange or biofilm distribution in relation to three-dimensional geometry.
Saliva as a moving film
Whole saliva is produced at rates that vary widely with stimulation, and its composition and functions have been reviewed [14, 15]. Its distribution in the mouth is as a thin film. From the surface area of the mouth and the volume of saliva present before and after swallowing, Collins and Dawes calculated an average film thickness on the order of 0.1 mm [16]. The film moves slowly and unevenly: experiments and calculations by Dawes and colleagues estimated film velocities that differ several-fold between oral sites and rise with stimulation, and showed that clearance of a diffusant from a surface depends on film velocity [17]. The exact velocity ranges should be taken from the full report; the order of magnitude for unstimulated flow is millimeters per minute.
Saliva is not a Newtonian fluid. Its viscosity falls with shear rate, it is viscoelastic, and both properties depend on the gland of origin and on the method of stimulation [18–20]. Stokes and Davies emphasized that characterizing saliva by viscosity at one shear rate is inadequate [20]. Any simulation must therefore adopt a rheological model with parameters taken from such measurements and must show how sensitive its conclusions are to them.
Hydrodynamics and oral biofilm
Flow influences biofilm in several ways. It delivers cells and nutrients and removes metabolic products; it applies shear that can detach cells; and the shear under which a biofilm grows conditions its structure and cohesive strength [21, 22]. Convective and diffusive transport of substrate shape biofilm architecture in models [23]. Oral biofilms grown at higher shear rates were stronger and harder to remove [24], and early flow-cell work related shear stress to adhesion and retention of oral microorganisms from saliva [25]. Dynamic in vitro biofilm models and printed microfluidic flow cells for microscopy are available [26, 27]. Computational fluid dynamics has been used to compute the wall shear stress by which high-velocity microdrops remove interproximal biofilm [28]. Surface roughness and free energy affect early biofilm formation and are the best-studied material determinants of plaque retention on restorations and abutments [29, 30].
The ecological view of dental disease holds that the local environment selects the composition of the biofilm: where clearance is poor, fermentation products and inflammatory exudate accumulate, pH and redox potential shift, and the community changes toward disease-associated species [31, 32]. On this view a niche with poor fluid exchange is not merely a place where more biofilm accumulates. It is a different habitat.
Fluid dynamics applied to salivary flow
The orthodontic simulations cited above treated the tooth and bracket as rigid boundaries and imposed a salivary flow over them [11, 12]. Simulations have also addressed salivary flow within ducts in the presence of stones [33]. These studies establish feasibility and show that predicted low-flow regions correspond to clinical sites of plaque accumulation around brackets. Their modeling choices, including the treatment of saliva as a bulk flow over the tooth, may not carry over to the confined geometry beneath an implant crown, where the fluid layer is thin compared with the dimensions of the niche.
Summary
It is established that restoration contour is associated with peri-implantitis, that saliva forms a thin, slowly moving, non-Newtonian film, and that hydrodynamics shapes biofilm. Computational fluid dynamics has linked appliance geometry to plaque retention in orthodontics. The fluid mechanics of the space between an implant restoration and the mucosa has not been characterized.
3. Research Gap
Four gaps are specific. The flow regime of saliva in the sub-contour and embrasure spaces of implant restorations has not been defined, so it is not known whether inertia, shear, or diffusion governs exchange there. No geometry-dependent, quantitative measure of fluid exchange exists for restorations. No experiment has compared predicted exchange with measured biofilm distribution on restoration replicas. And the clinical evidence cannot separate the effect of emergence angle from the effects of convexity and embrasure form, because angle is the only quantity that has been measured.
4. Research Objective and Central Hypothesis
The objective is to develop a validated method for computing fluid exchange around implant restorations from their geometry and to test whether it predicts biofilm retention.
The central hypothesis is that, among restorations with the same emergence angle, differences in profile convexity and embrasure geometry produce differences in the volume and surface area of poorly exchanged fluid that are large relative to numerical uncertainty; that the spatial distribution of biofilm on replicas in a flow cell correlates with the predicted distribution of fluid age; and that a retention index derived from the simulation explains more of the variance in biofilm accumulation than emergence angle does.
Secondary hypotheses are that the ranking of designs by retention index is robust to plausible variation in saliva rheology and flow rate; that predicted wall shear stress under resting flow is too low to account for differences in biofilm removal; and that at the clinical level the index is associated with site-specific plaque and bleeding after adjustment for hygiene access.
The central hypothesis is refuted if biofilm distribution on replicas is unrelated to predicted fluid age, or if emergence angle alone predicts accumulation as well as the index.
5. Scientific and Theoretical Framework
Flow regime
Consider a salivary film of thickness h moving at mean velocity U over a restoration. With density ρ and viscosity μ, the Reynolds number is
Re=(ρ U h)/μ (1)
Taking h≈10^(−4) m [16], U of the order of 10^(−5) to 10^(−4) m/s (millimeters per minute, the unstimulated order of magnitude [17]), and a viscosity no lower than that of water, Re is of order 10^(−3) to 10^(−2). Inertia is negligible. The flow is governed by the Stokes equations,
∇p=∇⋅(2 μ(γ) D), ∇⋅u=0 (2)
with velocity u, pressure p, rate-of-deformation tensor D, and a shear-rate-dependent viscosity μ(γ) to be specified from rheological data [19, 20].
Two consequences follow. First, the recirculation zones that appear in simulations at higher Reynolds number are not generated by inertia here. Recirculation does occur in creeping flow, as a sequence of progressively weaker eddies in corners narrower than a critical angle [34], and the re-entrant corner between a convex restoration and the mucosa is such a geometry. Fluid in corner eddies is isolated from the main stream and exchanges with it only by diffusion. Second, the wall shear stress is small. For a film sheared across its thickness,
τ_w≈(μ U)/h (3)
which for the values above and water-like viscosity is of order 10^(−4) to 10^(−3) Pa. Shear rates used to grow and to challenge oral biofilms in vitro span 0.1 to 50 s⁻¹ [24], corresponding to stresses up to tens of millipascals in water-like media, and removal of established biofilm requires far more [28]. Resting salivary flow should therefore not be expected to shear biofilm from a restoration, whatever its contour. Its role is transport.
Exchange by advection and diffusion
For a species with diffusivity D (small solutes, of order 10^(−9) m²/s in water; macromolecules and bacteria, orders of magnitude lower), concentration c obeys
(∂c)/(∂t)+u⋅∇c=D ∇^2c+S (4)
with source or sink S at the biofilm. The Péclet number Pe=UL/D, with L a length scale of the restoration, compares advection with diffusion. For L of a millimeter and small solutes, Pe is of order ten in the open film: advection dominates where fluid moves. In a sheltered niche where u→0, diffusion is the only mechanism, and the time to exchange across a niche of depth d is
t_D≈(d^2)/D (5)
Fluid age and the retention index
The age of fluid a(x), the mean time since the fluid at x entered the domain, is obtained by solving a steady transport equation with unit source,
u⋅∇a=D_a∇^2a+1 (6)
with a=0 at the inflow. Evaluated at the restoration and mucosal surfaces, a maps where the film is fresh and where it is stale. Two summary measures are proposed. The sheltered volume V_s is the volume of fluid whose age exceeds a multiple m of the mean transit time across the restoration. The retention index is the area-weighted mean of normalized surface age over the sub-contour surface Σ of the restoration, from the mucosal margin to the height of contour:
RI=1/(area(Σ))∫_Σ^ (a(x))/(a_(ref)) dA (7)
where a_(ref) is the surface age on a reference geometry (a straight profile at 20° emergence) under the same flow. RI equals one for the reference and rises as exchange worsens. Both measures depend on the choice of D_a; they are reported for a small-solute value and for a value representative of suspended bacteria.
From exchange to biofilm
A minimal local model for biofilm biomass B per unit area is
dB/dt=k_d c_b+μ_g(c_n,c_m) B−k_r(τ_f) B (8)
where the first term is deposition from suspended cells at concentration c_b, the second is growth at a rate that depends on nutrient concentration c_n and inhibitory metabolite concentration c_m, and the third is removal at a rate that depends on functional shear τ_f from tongue, cheek, bolus, and hygiene, which is much larger than resting salivary shear and is not computed by the resting-flow simulation. Geometry enters each term: it limits delivery and clearance through equations (4) and (5), and it shelters the surface from functional shear. The framework treats the fluid-exchange measures as predictors of the first two terms and leaves the third to the complementary problem of mechanical access. No parameter values for equation (8) are asserted; the equation organizes the hypotheses and is not used for quantitative prediction in the initial program.
6. Methodology
Stage 1: parametric geometry and simulation
Geometry. A single posterior implant crown between adjacent teeth on a mucosal base, parameterized by emergence angle (10° to 50°), profile shape (concave, straight, convex, with convexity quantified by the maximum deviation of the profile from the straight line joining platform and height of contour), circumferential variation, gingival embrasure height and width, and contact point position. A second family represents the intaglio of a full-arch prosthesis segment, parameterized by clearance from the mucosa and surface form (convex, flat, concave), since concave intaglio areas showed more biofilm accumulation in a clinical study [35]. Geometries are defined so that angle and convexity vary independently.
Fluid domain and boundary conditions. The film occupies the space between the restoration and mucosa and extends over adjacent surfaces. Resting flow is driven by a prescribed flux corresponding to film velocities across the published unstimulated range. Functional flow is represented by transient boundary motion: a moving wall for cheek or tongue and a pulsed flux for swallowing. Surfaces are no-slip. The free surface of the film is treated as a flat, shear-free boundary in the first instance; sensitivity to this simplification is examined in a subset with a deformable interface.
Rheology. A shear-thinning model with parameters fitted to published whole-saliva data for unstimulated and stimulated conditions [20], with Newtonian water-like and high-viscosity cases as bounds. Viscoelasticity is neglected initially and noted as a limitation.
Numerical method and verification. Finite-volume or finite-element solution of equations (2), (4), and (6). Verification includes comparison with the analytic corner-eddy solution for a wedge [34], a grid-refinement study with reported discretization error for V_s and RI [36], and a check that results are independent of domain extent. The well-documented variability of computational fluid dynamics results between groups working on the same problem [37] is the reason for depositing meshes and solver settings. Credibility assessment follows ASME V&V 40 and the corresponding FDA guidance [38, 39].
Outputs. For each geometry: maps of fluid age and wall shear; V_s; RI; depth and aspect ratio of any closed-streamline region. A global sensitivity analysis apportions variance in RI among geometric and fluid parameters.
Stage 2: flow-cell validation
Replicas. A subset of at least twelve geometries spanning the range of predicted RI, including pairs matched for emergence angle and differing in convexity, printed or milled in a restorative material and polished to a standardized roughness below the threshold reported to influence plaque formation [29, 30], mounted on a silicone mucosal analog in a transparent flow cell whose channel reproduces the simulated film thickness.
Biofilm. A defined multispecies consortium of early and later colonizers, or a saliva-derived microcosm, in a dynamic model [26], with artificial saliva of matched rheology perfused at flow rates corresponding to the simulated conditions. Cultures are run for 24, 48, and 96 hours.
Measurements. Confocal microscopy with viability staining gives biovolume and thickness maps registered to the replica geometry; total biomass is quantified by molecular methods. A non-reactive tracer washout experiment in each replica, without biofilm, measures residence time directly and validates the fluid-age prediction independently of biology.
Analysis. The primary analysis is the spatial correlation between predicted surface age and measured biovolume across surface patches, with replica as a random effect. The secondary analysis regresses total sub-contour biomass on RI and on emergence angle and compares the models. For angle-matched pairs, the difference in biomass is tested directly. Each condition is replicated in at least three independent runs.
Stage 3: clinical association
Design. Cross-sectional study of single posterior implant crowns for which the design file or an intraoral scan is available, reported to STROBE [40].
Exposure. RI computed from the three-dimensional geometry, with emergence angle and convexity measured on the same model, removing the limitation of two-dimensional radiographic measurement [3].
Outcomes. Site-level plaque by disclosing and standardized photography, bleeding on probing, probing depth, and peri-implant crevicular fluid volume. The study is not powered for peri-implantitis.
Covariates. A measure of hygiene access, self-reported interdental cleaning, keratinized mucosa, restoration material, time in function, and patient-level factors.
Analysis. Mixed models with patient as a random effect; the hypothesis is that RI is associated with plaque after adjustment for access. Sample size is to be set after Stage 2 from the observed relation between RI and biomass and from pilot variance of site-level plaque; none is assumed.
Bias, confounding, and reproducibility
In Stage 2, replicas differ in geometry only; material, roughness, and inoculum are held constant, and image analysis is automated and blinded. In Stage 3, contour is confounded with implant position, tissue thickness, and access; the association is interpreted as hypothesis-supporting only. All geometries, meshes, solver settings, and analysis code are to be made available.
Ethics
Stage 3 requires institutional review board approval and consent. Saliva-derived inocula in Stage 2 require donor consent.
7. Analytical Development
Diffusive exchange time of a sub-contour niche
Equation (5) gives the scale. For a small solute with D≈10^(−9) m²/s, a niche 0.5 mm deep exchanges in about 4 minutes, one of 1 mm in about 17 minutes, and one of 2 mm in about 67 minutes. For a macromolecule or a suspended bacterium, with diffusivity two to four orders of magnitude lower, the corresponding times range from hours to far longer than the interval between meals. These values follow from equation (5) with assumed textbook diffusivities and are order-of-magnitude illustrations. They are not measurements.
The quadratic dependence is the important feature. Doubling the depth of the sheltered zone quadruples the time for acid produced in the biofilm to leave and for buffer and antimicrobial proteins to arrive. A sub-contour space one millimeter deep is ventilated between meals. One two or three millimeters deep is not. This offers a physical reason why the association with disease should strengthen sharply with the degree of over-contour, as the clinical data suggest for the combination of wide angle and convex profile [3].
How angle and convexity determine niche depth
In a two-dimensional section, let the restoration emerge from the platform at height zero and reach its height of contour at height H and horizontal offset R, so that the mean emergence angle from the long axis is θ=arctan(R/H). A straight profile leaves a wedge-shaped space between restoration and mucosa whose depth, measured from the outer opening to the margin, is of order R. A convex profile with maximum outward deviation Δ from the straight line moves the restoration surface toward the mucosa near the margin and narrows the opening of the wedge while its depth is unchanged; the aspect ratio of depth to opening increases. A concave profile does the reverse. For the same θ, then, the three profiles have similar niche depth and different opening widths.
In creeping flow over a cavity, penetration of the outer flow into the cavity decreases rapidly as the aspect ratio rises, and closed eddies occupy the deeper part [34]. The effective diffusive depth d in equation (5) is therefore close to zero for a concave profile, which the outer flow sweeps, and close to the full depth R for a convex profile. With R=Htanθ, and H of 2 to 3 mm for a bone-level implant, an angle of 30° gives R of 1.2 to 1.7 mm and an angle of 45° gives 2 to 3 mm. By equation (5) the corresponding exchange times for a convex profile differ by a factor of about three. This reasoning predicts an interaction between angle and convexity of the kind reported [3], and predicts that angle should matter little when the profile is concave. It also suggests why tissue-level implants, whose restorative margin lies at or above the mucosa, showed no association: the wedge is open to the mouth.
Stimulated flow and function
During stimulation, film velocity rises by one to two orders of magnitude [17]. The Reynolds number remains below one for the film, so the creeping-flow picture persists, and the main effect is a proportional reduction in transit time over exposed surfaces. Corner eddies are a geometric feature of creeping flow and do not wash out as velocity increases. A sheltered zone therefore remains sheltered during stimulated flow, and the contrast between exposed and sheltered surfaces grows. Transient, larger-scale motions such as swallowing, rinsing, and tongue sweeping reach higher Reynolds numbers and can flush wider openings. The simulation of functional flow in Stage 1 is intended to establish which geometries are flushed by such events and which are not.
Predictions
The analysis predicts that (a) wall shear under resting flow is orders of magnitude too low to differentiate designs by biofilm removal; (b) surface fluid age rises steeply with the aspect ratio of the sub-contour space; (c) convexity modifies the effect of angle; (d) tracer washout from replicas shows a slow tail whose time constant scales with the square of niche depth; and (e) biofilm in sheltered zones differs in composition as well as in amount, which the microcosm experiments can examine by sequencing.
8. Discussion
The proposal supplies a physical mechanism for an epidemiological association and a way to measure it. Its first product, the scale analysis, already constrains what kind of mechanism is plausible. The salivary film is too thin and too slow to exert meaningful shear on a restoration or to generate inertial vortices of the kind seen in simulations at higher Reynolds number. If contour affects biofilm through saliva, it does so by limiting exchange. Models that seek a critical wall shear stress under resting flow would be looking in the wrong place.
The orthodontic simulations reported vortices gingival to brackets [11]. Without the full text it cannot be determined the Reynolds number at which they were run or whether the reported structures are corner eddies of creeping flow or inertial recirculation under an imposed bulk flow. Either way, the correspondence between low-flow zones and plaque that those studies observed is consistent with an exchange-limited mechanism. The modeling assumptions should be compared in detail before the present approach is finalized.
Competing explanations for the contour association remain. Over-contoured restorations impede brushes and floss, and experiments with artificial biofilm on crown replicas show that contour affects removal by interdental devices [41, 42]. Over-contour also compresses the mucosa, displaces the margin subgingivally, and may alter the soft-tissue seal. Wide angles are more common when implants are placed deep or off-axis, which carry risks of their own. Fluid exchange is one pathway among these, and the flow-cell stage isolates it by removing instruments, tissue response, and implant position from the experiment. A positive result there shows that geometry is sufficient to redistribute biofilm through fluid exchange. It does not show how much of the clinical association that pathway explains.
Several modeling choices are open to challenge. The film is treated as occupying the sub-contour space fully, whereas in vivo the space contains crevicular fluid flowing outward from the sulcus, a second source that opposes inward transport and has been modeled around teeth. Mucosa is compliant and moves. The free surface is simplified. Saliva is viscoelastic and adsorbs to surfaces as a pellicle that changes boundary conditions for bacterial adhesion. Each of these could alter quantitative results; the question for validation is whether they alter rankings.
The retention index is a relative measure tied to a reference geometry and to chosen diffusivities, and its threshold for clinical concern is unknown. It should be presented as a comparative design metric and not as a risk estimate.
Finally, there is a possibility that the effect is real and unimportant. If functional flushing and hygiene dominate biofilm control at accessible sites, exchange under resting flow may matter only where access is already poor, in which case the index would add little to a measure of instrument access. The clinical stage includes an access covariate to examine this, and the two measures are best developed in parallel and compared.
9. Clinical Significance
If validated, the method would allow a restoration design to be evaluated for fluid exchange before it is made. Design software could display the sheltered region on the virtual restoration and report the index alongside conventional parameters, and the designer could reshape the transition (reducing convexity near the margin, opening the gingival embrasure, relieving a concave intaglio) and see the effect. The analysis suggests that convexity near the margin deserves more attention than the angle as such, and that for a given required flare a concave or straight transition that keeps the sub-contour space open to the outer film is preferable, which agrees with clinical recommendations reached on other grounds [5, 7].
For patients with reduced salivary flow, the framework predicts longer transit times everywhere and a larger penalty for sheltered geometry, which would argue for more conservative contours in that group. That inference is untested.
10. Limitations
Film thickness and velocity are taken from estimates for the mouth as a whole and for a few sites; values adjacent to implant restorations have not been measured. Rheological parameters vary between individuals and with stimulation. The simulations neglect viscoelasticity, crevicular fluid outflow, tissue compliance, and pellicle. The retention index depends on reference geometry and assumed diffusivities. Flow-cell biofilms differ from in vivo plaque in composition, nutrient supply, and host factors. The clinical stage is cross-sectional and confounded. The diffusion-time estimates are order-of-magnitude arithmetic with assumed diffusivities. The comparison with orthodontic simulations is limited by reliance on abstracts. The patent review was not a professional search.
11. Future Validation
In situ validation would follow the flow-cell stage: replicas of differing predicted index worn on removable appliances by volunteers, with biofilm quantified after fixed periods, would test the prediction under real saliva and function. A prospective cohort relating the index at delivery to incident mucositis would be needed before any claim about disease. Measurement of film velocity near restorations, by tracer clearance, would replace assumed boundary conditions. Independent groups should reproduce the simulations from the deposited geometries, given known inter-laboratory variability in computational fluid dynamics [37].
12. Conclusion
Restoration contour is associated with peri-implantitis, and the explanation has rested on an unexamined notion of plaque retention. A scale analysis shows that resting saliva around a restoration is a creeping film whose shear is negligible and whose effect on biofilm must be through transport, with exchange times that grow as the square of the depth of any sheltered space. A retention index based on simulated fluid age expresses this for a given geometry, and the analysis predicts an interaction between emergence angle and convexity like the one observed clinically. The simulations, flow-cell experiments, and clinical study proposed here would test whether the index predicts where biofilm is retained. None has been performed.
Cite this paper
Jain S, Gupta A. Salivary Microfluidics Around Implant Restorations: A Computational and Experimental Methodology for Relating Prosthetic Geometry to Fluid Exchange, Microbial Retention, and Local Disease Risk. Working paper. Center for Implant Dentistry; 2026. Available from: https://bayareaimplantdentistry.com/clinician-research-and-publications/salivary-microfluidics-implant-restorations
References
- Su H, Gonzalez-Martin O, Weisgold A, Lee E. Considerations of implant abutment and crown contour: critical contour and subcritical contour. Int J Periodontics Restorative Dent. 2010;30(4):335-43. PMID: 20664835 (opens in a new tab).
- Gomez-Meda R, Esquivel J, Blatz MB. The esthetic biological contour concept for implant restoration emergence profile design. J Esthet Restor Dent. 2021;33(1):173-184. doi:10.1111/jerd.12714 (opens in a new tab). PMID: 33470498 (opens in a new tab).
- Katafuchi M, Weinstein BF, Leroux BG, Chen YW, Daubert DM. Restoration contour is a risk indicator for peri-implantitis: A cross-sectional radiographic analysis. J Clin Periodontol. 2018;45(2):225-232. doi:10.1111/jcpe.12829 (opens in a new tab). PMID: 28985447 (opens in a new tab).
- Yi Y, Koo KT, Schwarz F, Ben Amara H, Heo SJ. Association of prosthetic features and peri-implantitis: A cross-sectional study. J Clin Periodontol. 2020;47(3):392-403. doi:10.1111/jcpe.13251 (opens in a new tab). PMID: 31912511 (opens in a new tab).
- Mattheos N, Janda M, Acharya A, Pekarski S, Larsson C. Impact of design elements of the implant supracrestal complex (ISC) on the risk of peri-implant mucositis and peri-implantitis: A critical review. Clin Oral Implants Res. 2021;32 Suppl 21:181-202. doi:10.1111/clr.13823 (opens in a new tab). PMID: 34642979 (opens in a new tab).
- Soulami S, Slot DE, van der Weijden F. Implant-abutment emergence angle and profile in relation to peri-implantitis: A systematic review. Clin Exp Dent Res. 2022;8(4):795-806. doi:10.1002/cre2.594 (opens in a new tab). PMID: 35713938 (opens in a new tab).
- Lin GH, Lee E, Barootchi S, Rosen PS, Curtis D, Kan J, et al. The influence of prosthetic designs on peri-implant bone loss: An AO/AAP systematic review and meta-analysis. J Periodontol. 2025;96(6):634-651. doi:10.1002/JPER.24-0144 (opens in a new tab). PMID: 40489290 (opens in a new tab).
- Hamilton A, Putra A, Nakapaksin P, Kamolroongwarakul P, Gallucci GO. Implant prosthodontic design as a predisposing or precipitating factor for peri-implant disease: A review. Clin Implant Dent Relat Res. 2023;25(4):710-722. doi:10.1111/cid.13183 (opens in a new tab). PMID: 36691784 (opens in a new tab).
- Pirc M, Esquivel J, Patrizzi A, Jung RE, Strauss FJ. Emergence Profile Angle Matters-Restoring Peri-Implant Health by Adjusting Prosthetics-A Narrative Review. J Esthet Restor Dent. 2026;38(6):1188-1196. doi:10.1111/jerd.70109 (opens in a new tab). PMID: 41618482 (opens in a new tab).
- Monje A, Pons R, Soldini MC, Rosen PS, Nart J, Buser D. Role of Local Factors on the Occurrence of Peri-Implantitis: A Cross-Sectional Analysis. Clin Oral Implants Res. 2025;36(12):1614-1625. doi:10.1111/clr.70036 (opens in a new tab). PMID: 40916656 (opens in a new tab).
- Zhu P, Lin H, Han Y, Lin Y, Xu Y, Zhang Z. A computational fluid dynamic analysis of peri-bracket salivary flow influencing the microbial and periodontal parameters. PLoS One. 2013;8(4):e62242. doi:10.1371/journal.pone.0062242 (opens in a new tab). PMID: 23620815 (opens in a new tab).
- Xue J, Zhu P, He Y, Li Q, Xu Y. Application of computational fluid dynamics models for the evaluation of salivary flow patterns and related bacterial accumulation around orthodontic brackets. Orthod Craniofac Res. 2020;23(3):291-299. doi:10.1111/ocr.12369 (opens in a new tab). PMID: 32011803 (opens in a new tab).
- Serino G, Ström C. Peri-implantitis in partially edentulous patients: association with inadequate plaque control. Clin Oral Implants Res. 2009;20(2):169-74. doi:10.1111/j.1600-0501.2008.01627.x (opens in a new tab). PMID: 19077152 (opens in a new tab).
- Humphrey SP, Williamson RT. A review of saliva: normal composition, flow, and function. J Prosthet Dent. 2001;85(2):162-9. doi:10.1067/mpr.2001.113778 (opens in a new tab). PMID: 11208206 (opens in a new tab).
- Dawes C. Physiological factors affecting salivary flow rate, oral sugar clearance, and the sensation of dry mouth in man. J Dent Res. 1987;66 Spec No:648-53. doi:10.1177/00220345870660S107 (opens in a new tab). PMID: 3476629 (opens in a new tab).
- Collins LM, Dawes C. The surface area of the adult human mouth and thickness of the salivary film covering the teeth and oral mucosa. J Dent Res. 1987;66(8):1300-2. doi:10.1177/00220345870660080201 (opens in a new tab). PMID: 3476596 (opens in a new tab).
- Dawes C, Watanabe S, Biglow-Lecomte P, Dibdin GH. Estimation of the velocity of the salivary film at some different locations in the mouth. J Dent Res. 1989;68(11):1479-82. doi:10.1177/00220345890680110201 (opens in a new tab). PMID: 2531176 (opens in a new tab).
- Schwarz WH. The rheology of saliva. J Dent Res. 1987;66 Spec No:660-6. doi:10.1177/00220345870660S109 (opens in a new tab). PMID: 3476631 (opens in a new tab).
- van der Reijden WA, Veerman EC, Amerongen AV. Shear rate dependent viscoelastic behavior of human glandular salivas. Biorheology. 1993;30(2):141-52. doi:10.3233/bir-1993-30205 (opens in a new tab). PMID: 8400152 (opens in a new tab).
- Stokes JR, Davies GA. Viscoelasticity of human whole saliva collected after acid and mechanical stimulation. Biorheology. 2007;44(3):141-60. PMID: 17851164 (opens in a new tab).
- Stoodley P, Sauer K, Davies DG, Costerton JW. Biofilms as complex differentiated communities. Annu Rev Microbiol. 2002;56:187-209. doi:10.1146/annurev.micro.56.012302.160705 (opens in a new tab). PMID: 12142477 (opens in a new tab).
- Busscher HJ, van der Mei HC. Microbial adhesion in flow displacement systems. Clin Microbiol Rev. 2006;19(1):127-41. doi:10.1128/CMR.19.1.127-141.2006 (opens in a new tab). PMID: 16418527 (opens in a new tab).
- Picioreanu C, Van Loosdrecht MC, Heijnen JJ. Effect of diffusive and convective substrate transport on biofilm structure formation: a two-dimensional modeling study. Biotechnol Bioeng. 2000;69(5):504-15. doi:10.1002/1097-0290(20000905)69:5<504::aid-bit5>3.0.co;2-s (opens in a new tab). PMID: 10898860 (opens in a new tab).
- Paramonova E, Kalmykowa OJ, van der Mei HC, Busscher HJ, Sharma PK. Impact of hydrodynamics on oral biofilm strength. J Dent Res. 2009;88(10):922-6. doi:10.1177/0022034509344569 (opens in a new tab). PMID: 19783800 (opens in a new tab).
- Christersson CE, Glantz PO, Baier RE. Role of temperature and shear forces on microbial detachment. Scand J Dent Res. 1988;96(2):91-8. doi:10.1111/j.1600-0722.1988.tb01413.x (opens in a new tab). PMID: 3162604 (opens in a new tab).
- Sánchez MC, Alonso-Español A, Ribeiro-Vidal H, Alonso B, Herrera D, Sanz M. Relevance of Biofilm Models in Periodontal Research: From Static to Dynamic Systems. Microorganisms. 2021;9(2). doi:10.3390/microorganisms9020428 (opens in a new tab). PMID: 33669562 (opens in a new tab).
- Kristensen MF, Leonhardt D, Neland MLB, Schlafer S. A 3D printed microfluidic flow-cell for microscopy analysis of in situ-grown biofilms. J Microbiol Methods. 2020;171:105876. doi:10.1016/j.mimet.2020.105876 (opens in a new tab). PMID: 32087186 (opens in a new tab).
- Rmaile A, Carugo D, Capretto L, Aspiras M, De Jager M, Ward M, et al. Removal of interproximal dental biofilms by high-velocity water microdrops. J Dent Res. 2014;93(1):68-73. doi:10.1177/0022034513510945 (opens in a new tab). PMID: 24170371 (opens in a new tab).
- Quirynen M, Bollen CM. The influence of surface roughness and surface-free energy on supra- and subgingival plaque formation in man. A review of the literature. J Clin Periodontol. 1995;22(1):1-14. doi:10.1111/j.1600-051x.1995.tb01765.x (opens in a new tab). PMID: 7706534 (opens in a new tab).
- Teughels W, Van Assche N, Sliepen I, Quirynen M. Effect of material characteristics and/or surface topography on biofilm development. Clin Oral Implants Res. 2006;17 Suppl 2:68-81. doi:10.1111/j.1600-0501.2006.01353.x (opens in a new tab). PMID: 16968383 (opens in a new tab).
- Marsh PD. Are dental diseases examples of ecological catastrophes?. Microbiology (Reading). 2003;149(Pt 2):279-294. doi:10.1099/mic.0.26082-0 (opens in a new tab). PMID: 12624191 (opens in a new tab).
- Kolenbrander PE, Palmer RJ Jr, Periasamy S, Jakubovics NS. Oral multispecies biofilm development and the key role of cell-cell distance. Nat Rev Microbiol. 2010;8(7):471-80. doi:10.1038/nrmicro2381 (opens in a new tab). PMID: 20514044 (opens in a new tab).
- Zhu P, Lin Y, Lin H, Xu Y, Zheng Q, Han Y. Computational fluid dynamics analysis of salivary flow and its effect on sialolithogenesis. Oral Dis. 2014;20(6):624-30. doi:10.1111/odi.12182 (opens in a new tab). PMID: 24164693 (opens in a new tab).
- Moffatt HK. Viscous and resistive eddies near a sharp corner. J. Fluid Mech.. 1964;18(1):1-18. doi:10.1017/s0022112064000015 (opens in a new tab).
- Able FB, Campanha NH, Younes IA, Sartori IAM. Evaluation of the intaglio surface shape of implant-supported complete-arch maxillary prostheses and its association with biological complications: An analytical cross-sectional study. J Prosthet Dent. 2022;128(2):174-180. doi:10.1016/j.prosdent.2020.12.028 (opens in a new tab). PMID: 33563467 (opens in a new tab).
- Roache PJ. Perspective: A Method for Uniform Reporting of Grid Refinement Studies. Journal of Fluids Engineering. 1994;116(3):405-413. doi:10.1115/1.2910291 (opens in a new tab).
- Steinman DA, Hoi Y, Fahy P, Morris L, Walsh MT, Aristokleous N, et al. Variability of computational fluid dynamics solutions for pressure and flow in a giant aneurysm: the ASME 2012 Summer Bioengineering Conference CFD Challenge. J Biomech Eng. 2013;135(2):021016. doi:10.1115/1.4023382 (opens in a new tab). PMID: 23445061 (opens in a new tab).
- American Society of Mechanical Engineers. Assessing credibility of computational modeling through verification and validation: application to medical devices. ASME V&V 40-2018. New York: ASME; 2018.
- U.S. Food and Drug Administration. Assessing the credibility of computational modeling and simulation in medical device submissions: guidance for industry and Food and Drug Administration staff. Silver Spring (MD): FDA; 2023.
- von Elm E, Altman DG, Egger M, Pocock SJ, Gøtzsche PC, Vandenbroucke JP. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Lancet. 2007;370(9596):1453-7. doi:10.1016/S0140-6736(07)61602-X (opens in a new tab). PMID: 18064739 (opens in a new tab).
- Tuna T, Kuhlmann L, Bishti S, Sirazitdinova E, Deserno T, Wolfart S. Removal of simulated biofilm at different implant crown designs with interproximal oral hygiene aids: An in vitro study. Clin Oral Implants Res. 2019;30(7):627-636. doi:10.1111/clr.13448 (opens in a new tab). PMID: 31033028 (opens in a new tab).
- Sirinirund B, Siqueira R, Li J, Mendonça G, Zalucha J, Wang HL. Effects of crown contour on artificial biofilm removal efficacy with interdental cleaning aids: An in vitro study. Clin Oral Implants Res. 2023;34(8):783-792. doi:10.1111/clr.14105 (opens in a new tab). PMID: 37269176 (opens in a new tab).
- Astra Tech Inc, applicant. Method and system for designing custom restorations for dental implants. European patent application EP 1 973 491 A1. Priority 2006 Jan 5; published 2008 Oct 1 (related CN 101400317 A).
- Dentsply Sirona Inc, applicant. Method, system and devices for instant automated design of a customized dental object. United States patent application publication US 2022/0296344 A1. Priority 2019 Sep 5; published 2022 Sep 22 (granted as US 12,465,469 B2; related WO 2021/046147 A1).
- Dental-abutment design system and method. United States patent application publication US 2020/0138551 A1. Priority 2018 Nov 6.